User manual
PSW Series Programming Manual
80
SYSTem:BEEPer[:IMMediate]
Set
Query
Description
This command causes an audible tone to be
generated by the instrument. The duration time is
specified in seconds.
Syntax
Query Syntax
SYSTem:BEEPer[:IMMediate]
{<NR1>|MINimum|MAXimum}
SYSTem:BEEPer[:IMMediate]? [MINimum|MAXimum]
Parameter
<NR1>
0 ~ 3600 seconds.
MINimum
Sets the beeper time to the minimum (0
seconds)
MAXimum
Sets the beeper time to the maximum
(3600 seconds)
Return parameter
<NR1>
Returns the remaining beeper duration
time in seconds or returns the maximum
or minimum beeper time in seconds (for
the [MINimum|MAXimum] query
parameters).
Example 1
SYST:BEEP 10
**after a 2 second wait**
SYST:BEEP?
>8
The first command turns the beeper on for 10
seconds. After 2 seconds the SYST:BEEP? query
returns the remaining beeper time (8 seconds).
Example 2
SYST:BEEP? MAX
>3600
Returns the maximum settable beeper time in
seconds.
SYSTem:CONFigure:BEEPer[:STATe]
Set
Query
Description
Sets or queries the buzzer state on/off.
Syntax
Query Syntax
SYSTem:CONFigure:BEEPer[:STATe] {OFF|ON|0|1}
SYSTem:CONFigure:BEEPer[:STATe]?