User manual

REMOTE CONTROL
107
Error Code
Description
-100 Command
Error
This is the generic syntax error for devices that
cannot detect more specific errors. This code
indicates only that a Command Error as
defined in IEEE 488.2,11.5.1.1.4 has occurred.
-102 Syntax error
An unrecognized command or data type was
encountered; for example, a string was received
when the device does not accept strings.
-103 Invalid
separator
The parser was expecting a separator and
encountered an illegal character; for example,
the semicolon was omitted after a program
message unit,
MEAS:VOLT:DC?:MEASCURR:DC?
-104 Data type
error
The parser recognized a data element different
than one allowed; for example, numeric or
string data was expected but block data was
encountered.
-108 Parameter
not allowed
More parameters were received than expected
for the header; for example, the KLOCk
command only accepts one parameter, so
receiving SYSTem:KLOCk 1,0 is not allowed.
-109 Missing
parameter
Fewer parameters were recieved than required
for the header; for example, the KLOCk
command requires one parameter, so receiving
KLOCk is not allowed.
-111 Header
separator error
A character which is not a legal header
separator was encountered while parsing the
header; for example, no white space followed
the header, thus APPL5,1 is an error.