User manual

REMOTE CONTROL
31
Configure GPIB Interface
To use GPIB, the optional GPIB to USB (GUG-001) adapter must be
used. The GPIB to USB adapter must be connected before the PSW
is turned on. Only one GPIB address can be used at a time.
Configure GPIB
1. Ensure the PSW is off before proceeding.
2. Connect the USB cable from the rear panel
USB B port on the PSW to the USB A port on
the GPIB to USB adapter.
3. Connect a GPIB cable from a GPIB controller to
the GPIB port on the adapter.
Type B plug for
PSW series
Type A plug
From
computer
GUG-001
4. Turn the PSW on.
5. Press the Function key to enter the
Normal configuration settings.
Page 21
Set the following GPIB settings:
F-22 = 1
Set the rear panel USB port to
GPIB-USB (GUG-001)
F-23 = 0~30
Set the GPIB address (0~30)
GPIB constraints
Maximum 15 devices altogether, 20m cable
length, 2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection