User manual

REMOTE CONTROL
59
Parameter
0
<NR1>Turns blink OFF
OFF
Turns blink OFF
1
<NR1> Turns blink ON
ON
Turns blink ON
Return parameter
0
<NR1>Turns blink OFF
1
<NR1>Turns blink ON
Example
DISP:BLIN 1
Turns blink ON.
Initiate Commands
INITiate[:IMMediate]:NAME ............................................. 59
INITiate[:IMMediate]:NAME
Set
Description
The INITiate command starts the TRANsient or
OUTPut trigger.
See the trigger commands on page 76 for usage
details.
Syntax
INITiate[:IMMediate]:NAME {TRANsient|OUTPut}
Parameter
TRANSient
Starts the TRANsient trigger.
OUTPut
Starts the OUTPut trigger.
Example
INITiate:NAME TRANient