REMOTE CONTROL
75
Syntax
Query Syntax
[SOURce:]VOLTage:SLEW:RISing {<NRf>|MIN|MAX}
[SOURce:]VOLTage:SLEW:RISing? [MIN|MAX]
Parameter/Return
<NRf>
0.01V/s~60.00V/s (PSW 30-XX)
0.1V/s~160.0V/s (PSW 80-XX)
0.1V/s~320.0V/s (PSW 160-XX)
0.1V/s~500.0V/s (PSW 250-XX)
1V/s~1600V/s (PSW 800-XX)
MIN
Minimum rising voltage slew rate.
MAX
Maximum rising voltage slew rate.
Example
SOUR:VOLT:SLEW:RIS MAX
Sets the rising voltage slew rate to its maximum.
[SOURce:]VOLTage:SLEW:FALLing
Set
Query
Description
Sets or queries the falling voltage slew rate. This is
only applicable for CV slew rate priority mode.
Syntax
[SOURce:]VOLTage:SLEW:FALLing {<NRf>|MIN|MAX}
Query Syntax
[SOURce:]VOLTage:SLEW:FALLing? [MIN|MAX]
Parameter/Return
<NRf>
0.01V/s~60.00V/s (PSW 30-XX)
0.1V/s~160.0V/s (PSW 80-XX)
0.1V/s~320.0V/s (PSW 160-XX)
0.1V/s~500.0V/s (PSW 250-XX)
1V/s~1600V/s (PSW 800-XX)
MIN
Minimum voltage falling slew rate.
MAX
Maximum voltage falling slew rate.
Example
SOUR:VOLT:SLEW:FALL MIN
Sets the falling voltage slew rate to its minimum.