SYSTem:CONFigure:OUTPut:PON[:STATe]
Set
Query
Description
Sets the unit to turn the output ON/OFF at power-
up. This setting is only applied after the unit has
been reset.
Syntax
SYSTem:CONFigure:OUTPut:PON[:STATe]
{OFF|ON|0|1}
Query Syntax
SYSTem:CONFigure:OUTPut:PON[:STATe]?
Parameter
0
Output off at power up
OFF
Output off at power up
1
Output on at power up
ON
Output on at power up
Return Parameter
0
Output off at power up
1
Output on at power up
SYSTem:COMMunicate:ENABle
Set
Query
Description
Enables/Disables LAN, GPIB or USB remote
interfaces as well as remote services (Sockets, Web
Server).
This setting is applied only after the unit is reset.
Syntax
Query Syntax
SYSTem:COMMunicate:ENABle <mode>,<interface>
SYSTem:COMMunicate:ENABle? <interface>
Parameter
<mode>
OFF
Turns the selected mode off.
0
Turns the selected mode off.
ON
Turns the selected mode on.
1
Turns the selected mode on.
<interface>
GPIB
Select GPIB
USB
Select USB
LAN
Select LAN
SOCKets
Select Sockets
WEB
Select the web server
Return Parameter
0
The selected mode is off.
1
The selected mode is on.