Instructions

COMMUNICATION INTERFACE
109
Type B plug for
PSW series
Type A plug
From
computer
GUG-001
4. Turn the PSW on.
5. Press the Function key to enter the
Normal configuration settings.
Page 84
Set the following GPIB settings
F-22 = 1
Set the rear panel USB port to
GPIB-USB (GUG-001)
F-23 = 0~30 Set the GPIB address (0~30)
GPIB constraints
Maximum 15 devices altogether, 20m cable
length, 2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection
Configure the Ethernet Interface
The Ethernet interface can be configured for a number of different
applications. Ethernet can be configured for basic remote control or
monitoring using a web server or a telnet connection.
The PSW series supports DHCP connections so the instrument can
be automatically connected to an existing network. Alternatively,
network settings can also be manually configured.
Background
The Ethernet interface is used for remote
control and monitoring over a network.
Ethernet
configuration
MAC Address
(display only)
LAN