Instructions

OPERATION
45
Set to C.V. Mode
When setting the power supply to constant voltage mode, a current
limit must also be set to determine the crossover point. When the
current exceeds the crossover point, the mode switches to C.C.
mode. For details about C.V. operation, see page 20. C.C. and C.V.
mode have two selectable slew rates: High Speed Priority and Slew
Rate Priority. High Speed Priority will use the fastest slew rate for
the instrument while Slew Rate Priority will use a user-configured
slew rate.
Background
Before setting the power supply to C.V. mode,
ensure:
The output is off.
The load is connected.
Steps
1. Press the Function key. The
Function key will light up.
2. Set F-03 (V-I Mode Slew Rate
Select) to 0 (CV High Speed
Priority) or 2 (CV Slew Rate
Priority) and save.
Page 84
F-03
0 = CV High Speed Priority
2 = CV Slew Rate Priority
3. If CV slew rate priority was chosen
as the operating mode, set F-04
(Rising Voltage Slew Rate) and the
F-05 (Falling Voltage Slew Rate)
and save.
Page 84
F-04
F-05
0.1V/s~60V/s (PSW 30-XX)
0.1V/s~160V/s (PSW 80-XX)