Instructions
CONFIGURATION
75
Configuration
Configuration of the PSW power supplies is divided into five
different configuration settings: Normal Function, USB/GPIB, LAN,
Power ON Configuration and Calibration Settings. Power ON
Configuration differs from the other settings in that the settings
used with Power ON Configuration settings can only be set during
power up. The other configuration settings can be changed when
the unit is already on. This prevents some important configuration
parameters from being changed inadvertently. Power On
Configuration settings are numbered F-90 to F-95 and the other
configuration settings are numbers F-00 to F-61.
Configuration Table
Please use the configuration settings listed below when applying
the configuration settings.
Normal Function
Settings Setting Setting Range
Output ON delay time F-01 0.00s~99.99s
Output OFF delay time F-02 0.00s~99.99s
V-I mode slew rate select
F-03
0 = CV high speed priority
1 = CC high speed priority
2 = CV slew rate priority
3 = CC slew rate priority
Rising voltage slew rate F-04
0.01V/s~60.00V/s (PSW 30-XX)
0.1V/s~160.0V/s (PSW 80-XX)
Falling voltage slew rate F-05
0.01V/s~60.00V/s (PSW 30-XX)
0.1V/s~160.0V/s (PSW 80-XX)
Rising current slew rate F-06
0.01A/s~72.00A/s (PSW 30-36)
0.1A/s~144.0A/s (PSW 30-72)
0.1A/s~216.0A/s (PSW 30-108)
0.01A/s~27.00A/s (PSW 80-13.5)
0.01A/s~54.00A/s (PSW 80-27)
0.01A/s~81.00A/s (PSW 80-40.5)