Instructions
OPERATION
55
Master-Slave Parallel Overview
Description
When connecting the PSW power supplies in
parallel, up to 3 units can be used in parallel
and all units must be of the same model.
When the units are used in parallel, a number
of precautions and limitations apply. Please
read this overview before operating the power
supplies in parallel.
Limitations
Display
Only the master unit will display the voltage
and current.
OVP/ OCP
The master unit can shut down slave units
when OVP/OCP is tripped on the master
unit (if the slave connector is wired for shut
down on alarm).
OVP/OCP can be independently tripped on
each slave unit, however the shutdown of
the power or output of the unit is disabled.
Only the alarm will be enabled.
Remote monitoring
Voltage monitoring (VMON) and current
monitoring (IMON) are only supported on
the master unit.
Remote Sense
Please see the remote sense chapter for
details, page 50.
Voltage controlled remote control
Voltage controlled remote control can only
be used with the master unit.