User Manual
OPERATION
49
Basic Operation
This section describes the basic operations required to operate the
power supply.
Setting OVP/OCP → from page 49
C.V. mode → from page 51
C.C. mode → from page 54
Display modes → page 57
Panel lock → page 58
Remote sensing → from page 59
Before operating the power supply, please see the Getting Started
chapter, page 9.
Setting OVP/OCP Levels
The OVP level has a selectable range of 10% to
110% of the rated output voltage. The OCP
level has a selectable range 10%~ 110% of the
rated output current, alternatively the OCP
level can also be turned off. The OVP and OCP
level is set to 110% by default.
When one of the protection measures are on,
ALM is shown on the panel display. By default,
the power switch will turn off when any of the
protection levels are tripped.
ALM
V
A