User Manual
OPERATION
71
Remote monitoring
Voltage monitoring (VMON) and current
monitoring (IMON) are only supported on
the master unit.
The VMON voltage represents the total
voltage of the all the serialized units.
Remote Sense
Please see the remote sense chapter for
details, page 59.
External Voltage and Resistance Control
Voltage/Resistance controlled remote
control can only be used with the master
unit.
The full scale voltage (in series) is equivalent
to the maximum external voltage or
resistance.
Slew Rate
The actual slew rate is double that of the
setting slew rate. I.e., A slew rate setting of
60.00V/s is actually 120V/s when in series.
Internal Resistance
The internal resistance is actually twice that
of the setting value.
Bleeder Control
The Master unit is used to control the
bleeder settings. The bleeder resistor is
always turned on for the slave unit in series
mode.