When setting the OVP for
channel3, select CH3 meter
using the CH1/CH3 meter
switch.
CH1
CH3
5. When finished, slide the
rear panel switch to the
“Normal” position. The
OVP indicator on the front
panel turns Off.
CH3CH2CH1
O.V.P . SETNORMAL
O
.V.
P
.
SET
TRIP
O.V.P.
SET
TRIP
When OVP is
activated....
The OVP activates when one
of channel1/2/3 output
voltage exceeds the OVP
settin
g
. The indicator turns red
(tripped), and the output is
shut Off immediately.
O
.V.P.
SET
TRIP
O.V.P.
SET
TRIP
OUTPUT
ON / OFF
OUTPUT
ON / OFF
SPD-3606 User Manual
24
Load Cable Connection
Standard
accessory
(GTL-104)
1. Turn the terminal
counterclockwise and
loose the screw.
2. Insert the cable
terminal.
C
H
2
G
N
D
0
6
0
V
,
3
A
0
3
0
V
,
6
A
2
1
3. Turn the terminal
clockwise and tighten
the screw.
C
H
2
G
N
D
0
6
0
V
,
3
A
0
3
0
V
,
6
A
3
Banana plug
Insert the plug into the
socket.
C
H
2
G
N
D
0
6
0
V
,
3
A
0
3
0
V
,
6
A
Wire type
When using load cables other than the attached,
make sure they have enough current capacity for
minimizing cable loss and load line impedance.
Voltage drop across a wire should not excess 0.5V.
The following list is the wire current rating at
450A/cm
2
.
Wire size (AWG) Maximum current (A)
20 2.5
18 4
16 6
14 10
12 16