User guide

PERFORMANCE VERIFICATION
41
Default SPD-3606 setting
The following is the required front and rear panel setting before
running each verification.
Range
60V/3A
60V/3A
30V/6A
30V/6A
60V/3A
30V/6A
Output Voltage
Tracking Series
OVP
Output Current
Tracking Parallel
OVP setting SW
Normal position
CH3CH2CH1
O.V.P. SETNORMAL
Channel1/3 meter
SW
CH1 position
CH1
CH3
Tracking SW
Independent position,
+ .
T
R
A
C
K
I
N
G
I
N
D
E
P
.
S
E
R
I
E
S
P
A
R
A
L
L
E
L
Channel1/2/3
Voltage knob
Minimum position
V
O
L
T
A
GE
Channel1/2
Current knob
Minimum position
C
U
R
R
E
N
T
SPD-3606 User Manual
42
Output Voltage Verification
Check item
Minimum output Voltage accuracy
Maximum output Voltage accuracy
Voltage meter accuracy (Output On)
Voltage meter accuracy (Output Off)
Connection
Digital Multimeter
SPD-3606
V
A
Verification step
1. Set the SPD panel according to the Default
setting list, page41.
2. Connect SPD Channel 1 and
Digital Multimeter Voltage
terminal.
CH1
0 60V , 3A
0 30V , 6A
3. Power up SPD and Digital Multimeter.
4. Turn up the SPD Current knob to
the maximum.
C
U
R
R
E
N
T