PERFORMANCE VERIFICATION
45
Tracking Series Voltage Verification
Check item
• Minimum output Voltage accuracy
• Maximum output Voltage accuracy
Connection
Digital Multimeter
SPD-3606
V
A
Verification step
1. Set the SPD panel according to the Default setting
list, page41, except for the tracking switch (see
below).
2. Set the tracking switch position
to Series,
+ .
T
R
A
C
K
I
N
G
I
N
D
E
P
.
S
E
R
I
E
S
P
A
RA
L
L
E
L
3. Connect SPD Channel 1 and
Digital Multimeter Voltage
terminal.
CH1
0 60V , 3A
0 30V , 6A
4. Power up SPD and Digital Multimeter.
SPD-3606 User Manual
46
5. Turn up the SPD Current
knob, both Channel1 and
Channel2, to the maximum.
C
U
R
R
E
N
T
6. Turn On the SPD output.
O
U
T
P
U
T
ON / OFF
O
U
T
P
U
T
ON / OFF
7. Turn up the SPD Channel1
Voltage knob until the
Multimeter reading shows
1.000V.
V
O
L
T
A
GE
(1.000V)
Minimum
tracking series
output voltage
8. Connect the Multimeter to
SPD Channel2 and record
the reading. Here is the
acceptance range.
CH2
0 60V , 3A
0
30
V
, 6
A
Channel2
0.985V ~ 1.015V
9. Connect Digital Multimeter
back to SPD Channel1.
CH1
0 60V , 3A
0 30V , 6A
10. Turn up the SPD Channel1
Voltage knob until the
Multimeter reading shows
60.00V.
V
O
L
T
A
GE
(60.00V)
Maximum
tracking series
output voltage
11. Connect the Multimeter to
SPD Channel2 and record
the reading. Here is the
acceptance range.
CH2
0 60V , 3A
0
30
V
, 6
A
Channel2
59.69V ~ 60.31V