User Manual

-
25
-
Guidance and manufacture’s declaration electromagnetic immunity
TC200 is intended for use in the electromagnetic environment specified below. The customer or the user of TC200 should assure that
they are used in such an environment.
Immunity test
IEC/EN 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
When the relative humidity is approximately
5% and there are synthetic materials
present, static charges approaching 15kV
can be generated.
Electrical fast transient/burst
IEC/EN 61000-4-4
±2 kV for power supply
lines
N/A
N/A
Surge
IEC/EN 61000-4-5
±1 kV for line to line
±2 kV for line to ground
N/A
N/A
Power frequency
(50/60Hz)
magnetic field
IEC/EN 61000-4-8
30 A/m
30 A/m
This test level assumes a minimum distance
between TC200 and sources of power
frequency magnetic field of at least 15 cm.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC/EN 61000-4-11
<5 % U
T
>95% dip in
U
T
for 0.5 cycle
40 % U
T
60% dip in U
T
for 5 cycles
70 % U
T
30% dip in U
T
for 25 cycles
<5 % U
T
>95% dip in
U
T
for 5 s
N/A
N/A
NOTE U
T
is the a.c. mains voltage prior to application of the test level.