Use Instructions

ELST026-AUM001-V0.6.2 Page 11 of 68
1.2 Safety instructions
1.2.1 General
Only qualified people who are trained in the application of the therapy may use the device.
For optimum treatment, an anamnesis must first be performed. Based on the find-
ings, a treatment plan with objectives will be formulated. Follow the treatment plan
during the therapy. This will limit possible risks, related to the treatment, to a mini-
mum.
Only treat patients with electrical implants (pacemaker) after obtaining medical advice.
Only a technician authorized by GymnaUniphy N.V. may open the equipment or the
accessories.
Always keep these user instructions with the equipment.
Follow the instructions and directions in these user instructions.
Only make use of standard and optional accessories listed in these instructions for use.
When the mains adapter is used then do not place the Acure or other equipment in such a
way the adapter cannot be easily removed from the mains outlet socket.
Place the equipment on a horizontal and stable surface.
Keep the ventilation openings at the bottom and rear of the equipment free.
Do not place the equipment in the sun or above a heat source.
Do not place any objects on the equipment.
Do not use the equipment in a damp area.
Do not let any liquid flow into the equipment.
Do not sterilize the equipment or accessories.
For cleaning and disinfecting the equipment and accessories: see § 5.2
The Medical Devices Directive 93/42/EEG from the European Commission requires devices
to be safe during their service life. Therefore, it is strongly recommended to perform at
least each two years a technical safety inspection. See § 5.1.1
1.2.2 Electrical safety
Only use the equipment in an area with facilities that meet the applicable legal regulations.
It is strictly forbidden using any external power supply or battery pack that is not specified
in this user manual. See § 7.6
Connect the medical grade power supply to an outlet that meets the locally applicable
requirements for medical areas. The device is Class II.
The output connectors are exclusively for use with the supplied cable of the hand piece and
the 1-pole patient cable.
1.2.3 Prevention of explosion
Do not use the equipment in an area where combustible gases or vapours are present.
Switch of the device before disinfecting or decontaminating the room where it is placed.
Switch off the equipment when it is not used.