Use Instructions

ELST026-AUM001-V0.6.2 Page 15 of 68
2 INSTALLATION
2.1 Receipt
Procedure
1. Check that the equipment did not get damaged during transport.
2. Check that the accessories are intact and complete. See § 7.6
3. Inform your supplier of any damage or defects by no later than within 3 working days after
receipt. Report the damage by telephone, fax, e-mail or letter.
Warning:
Do not use the equipment if it is damaged or defective.
2.2 Placement of the device
Procedure
1. Place the equipment on a horizontal and stable base.
Warning:
Keep the ventilation openings at the bottom and rear of the equipment free.
Do not place the equipment in the sun or above a heat source.
Do not use the equipment in a wet area.
2.3 Connection to the mains supply
Connect the device to the supply mains with the standard delivered external medical grade
power supply. One can use of a power adaptor socket when necessary. See § 7.6
Connect to a mains outlet that meets the locally applicable requirements for medical areas.
The power supply is suited for a nominal mains voltage of 100-240 V / 50-60 Hz.
2.4 Mains power or battery operation
The device has an internal Li-ion battery. It is strictly recommended to fully charge the bat-
tery before taking into use. The device can operate on the supply mains or on the battery
while performing the treatment.
The device has a battery management system that controls charging when connected to the
supply mains and warns when the battery gets exhausted.
The actual condition of the battery is visualized in the right upper corner of the LCD-screen.
Either;
Device is connected to the supply mains. (lightning sign in battery)
The treatment will be performed while making use of the mains power.
The battery management system takes care of recharging the battery.
Device is disconnected from the supply mains.
The treatment will be performed running on the battery.
The battery management system takes care that the treatment can only start if the battery
is sufficiently charged. (level >10 %)
The remaining battery charge is too low to start a treatment (level 10 %)