Use Instructions

ELST026-AUM001-V0.6.2 Page 16 of 68
There’s an extra indicator LED, near the Power On button at the backside of the device, if for
some reason the condition of the battery is not visible on the LCD.
No light
Device is powered Off
White full
Device is powered On. But not connected to the mains power
Green glowing
Device battery is charging
Green full
Device battery is fully charged 100%
Orange full
Device battery level is < 10%
Attention:
It is recommended to power the device by the supply mains whenever possible. This
will increase the service life of the battery.
2.5 Touchscreen application /save energy modes
The device has a touchscreen. All settings and treatment possibilities can be selected by touching
the applicable button.
Attention:
To save energy this device goes in sleep modes. Bring the device out of sleep modes by
just touching the touchscreen. (See § 4.7.1 standby time)
If theres no response or if the display does not light up, connect the device to the medical
grade power supply and switch on the device via the Power On button at the backside.
2.6 A functional device test is automatically performed
every time the device is switched on with the Power On button
If the self-test was successfully, the device enters the application screen.
If an error was encountered, an error message appears to guide the user on how to proceed.
2.7 Check proper functioning
Observe whether the device is operating normally, especially when other electrical or elec-
tronic devices or transmitters are operating nearby. In case the behavior is erratic, try reori-
enting or repositioning the device to mitigate possible electromagnetic disturbances. See
also §8.1.1.
2.8 Shutting down procedure
Press the stop button when the equipment is still generating current.
Discard the needle in the safety container when still in the hand piece.
Press the power off button at the back above the socket for the adapter.
Remove the adapter from the mains outlet socket if charging the battery is not necessary.