Use Instructions

ELST026-AUM001-V0.6.2 Page 24 of 68
4.
From now on are the and buttons on both the hand piece and display enabled to
adjust the intensity.
5.
Every single push on the or changes the intensity by one step, and is made visible
on the hand piece by a short blinking of the green illuminated ring.
The output window and value bars:
1.
The intensity value bar shows the actual
intensity value as a percentage of the maximum
allowed intensity value. (e.g. 300/2500)
2.
During the therapy increases the effective
treatment timer every second while the initial
preset treatment timer stays unaltered.
3.
The time value bar shows the effective
treatment time as a percentage of the set
treatment time. (e.g. 1:40/2:00)
4.
The administered electrical charge Q in milli-
Coulomb (mC) is updated once per second.
Attention:
There will be cases that for a certain treatment the charge to administer is more
or less predictable. In that case will the user be advised via a message on the
touchscreen (e.g. 30 mC level reached).
Additional the continuous green coloured illuminated ring on the hand piece will
shortly interrupt by an orange flashing light.
Nevertheless, the user can still decide to proceed the treatment based on the
evaluation of the echographic image.
Stop the therapy
Procedure:
1.
Once the user decides the administered electrical charge is sufficient, based on the
evaluation of the echographic image, the treatment must be stopped.
2.
Press the Start/Stop button on the hand piece . The intensity will fade out to zero.
3.
All parameter settings are retained.
4.
A Stop trigger on the hand piece can be seen as a kind of pause. Treatment can be
continued by pressing again the start button on the hand piece.
Attention:
5.
Pressing the stop button on the touchscreen ends the treatment. The parameters have
to be confirmed again on the touchscreen before the treatment can be resumed.
6.
This confirmation enables the Start/Stop button on the hand piece again.