Use Instructions

ELST026-AUM001-V0.6.2 Page 30 of 68
4.1.2.1 Performing the therapy in advanced modes
Beneath describes the necessary actions that need to be taken in correct order on both the device
and the hand piece to allow performing the treatment application.
For the visual colour feedback of the illuminated ring on the hand piece see ยง 4.4
Set a therapy
Procedure:
1.
Press on the corresponding petal to get direct
access for changing the parameter values of:
a.
Intensity step size
b.
Ramp curve
c.
Preset intensity as the target value
2.
Change treatment time if necessary with
the enabled and buttons
3.
Push the confirm button if you agree with all
the presets.
4.
This confirmation enables the Start button on
the hand piece. See content.
5.
The green illuminated ring on the hand piece flashes very briefly.
Start and perform a therapy
Procedure:
1.
The hand piece and its buttons function as a wired remote control
2.
Press the Start button on the hand piece
Attention:
The intensity will increase (fade-in) automatic to its preset value from the
moment that the Start button on the hand piece is pressed.
3.
During the fade-in time flashes the green
illuminated ring on the hand piece, while it
lights up continuously as soon as the preset
intensity value is reached.
4.
From the moment that the Preset intensity
value is reached can the intensity be fine-
tuned with the enabled and buttons
on either the hand piece or the touchscreen