Use Instructions

ELST026-AUM001-V0.6.2 Page 31 of 68
5.
Every single trigger on the and intensity
buttons is noticeable via a short blinking of the
green illuminated ring.
The output window and value bars:
6.
The intensity value bar in a circle shows the actual
intensity value as a percentage of the maximum
allowed intensity value. (e.g. 550/2500)
7.
During the therapy increases the effective
treatment timer every second while the initial
preset treatment timer stays unaltered.
8.
The time value bar in a circle shows the effective
treatment time as a percentage of the set
treatment time. (e.g. 0:08/2:00)
9.
Every second becomes the administered electrical
charge Q in milli-Coulomb (mC) updated.
Attention:
There will be cases that for a certain treatment application the administering
charge level is somehow nearly predictable. In that case will the user be made
attentive via a message on the touchscreen (e.g. 30 mC level reached).
Additional the continuous green coloured illuminated ring on the hand piece will
shortly interrupt by an orange flashing light.
Nevertheless, the user can still decide to proceed the application based on the
evaluation of the echographic image.
Stop the therapy
Procedure:
1.
Once the user agrees the administered electrical charge based on the evaluation of the
echographic image, the treatment must be stopped.
2.
Press the Stop button on the hand piece . The intensity will fade out to zero.
3.
All parameter settings are retained.
4.
A Stop trigger on the hand piece can be seen as a kind of pause. Treatment can be
continued by pressing again the start button on the hand piece.
Attention:
The intensity will increase (fade-in) automatic to its preset value from the
moment that the Start button on the hand piece is pressed.
5.
Pressing the stop button on the touchscreen ends the treatment. The parameters have
to be confirmed again on the touchscreen before the treatment can be resumed.
6.
This confirmation enables the Start/Stop button on the hand piece again.