Use Instructions

ELST026-AUM001-V0.6.2 Page 37 of 68
4.4 The coloured illuminated ring on the hand piece
The illuminated rings on the hand piece provides the following information.
The colours can be green (normal operation), orange (point to be attentive), red (warning) which
points that an immediate response of the operator is required.
Indication light
Situation
no light
hand piece is NOT yet in operate modes
or hand piece is no longer detected by the device
no reactions on any key presses
green
very briefly flashing
parameter settings are confirmed
hand piece is detected by the device and in operate modes
waiting state for the Start button to be pressed
green
blinking
hand piece operates in ‘advanced modes’
Start button is pressed
the light blinks during the ramp up-time until the Set intensity
value is reached; or;
the light blinks during the ramp down-time until the intensity
is zero.
green
continuous light
Hand piece is in operate modes
patient is currently treated with the Set Intensity
green
short interruption
Hand piece is in operate modes
each press on the intensity up or intensity down button is
short visible
green
slowly switching by
orange
Hand piece is in operate modes
signal to be attentive for the applied charge, although user
can decide to proceed the action without harm
orange
continuous light
Hand piece is just stopped delivering the set intensity
there is an open patient circuit detected during the therapy
green
rapidly switching by
orange
Hand piece is in operate modes
signal to warn the user for the actual situation. It is advised
to thoroughly check this situation, and finish the treatment as
soon as possible.
red
continuous light
hand piece is still detected by the device
immediate response from operator required
points to a failure modes
hand piece buttons do NOT react when pressed!