Use Instructions

ELST026-AUM001-V0.6.2 Page 44 of 68
Sound: Activate or deactivate the sound settings for specific situations.
they function as audio feedback to be attentive or as a warning.
they are not meant to function as alarm signals!
Parameter
range
default
Volume
1 .. 5
3
Disabled key pressed
On ; Off
On
Confirmation sound
On ; Off
On
Start button pushed
On ; Off
On
Stop button pushed
On ; Off
On
Open patient circuit detected
On ; Off
On
mC thresholds reached
On ; Off
On
Patient database use: default in On-modes
If the Patient database setting is switched to the Off-modes, then there is no possibility to save
the performed treatment sessions at all. And there is no add-sequence button available in the
options pane.
Additional the Patient database button in the left menu pane will disappear as long as this setting
is put in Off-modes.
System information: view the system information of the device.
Always have this information available when you contact the technical service department.
Hand piece:
Hand piece information: information of the hand piece. (act as read only screen)
Hand piece buttons , use default in On-modes
By default all 3 buttons on the hand piece are enabled.
But it is possible to switch this parameter to the Off-modes. In that case it is not possible to adjust
the set intensity by using the and buttons on the hand piece.
Hand piece test: allows to check the proper functionality of the buttons and the
coloured illuminated ring
Procedure to check the hand piece buttons
1.
Select: Hand piece test
2.
Follow the on screen instructions
3.
Press one hand piece button at the time
4.
The guided hand piece picture shows directly the corresponding result of the test activity
Attention: complete test performed of all 3 buttons, independent of certain system settings.