Use Instructions

ELST026-AUM001-V0.6.2 Page 47 of 68
Inspection result
1. A registration must be maintained of the technical safety inspections. Use the inspection report
in the appendix for this purpose. See § 8.2
2. Copy this appendix.
3. Complete the copied appendix.
4. Keep the inspection reports for at least 10 years.
The inspection is successful if all inspection items are passed.
Repair all faults on the equipment before the equipment is put back into operation.
By comparing the registered measurement values with previous measurements, a possible slowly-
deteriorating deviation can be ascertained.
5.2 Maintenance
Component
Check
Frequency
hand piece,
cable, connector
‘cleaning’
See § 5.2.3
after every treatment
electrode cable &
rubber electrode
‘cleaning’
See § 5.2.3 & § 5.2.4
after every treatment
electrode sponge
‘cleaning’
see § 5.2.5
after every treatment
fixation bandages
‘cleaning’
see § 5.2.6
at least 1x per week
main device
‘cleaning’’
see § 5.2.1 & § 5.2.2
at least 1x per week
acupuncture needle
‘disposable’
single use
dispose safely after
each use in special
container
Caution:
Accessories that come in contact with the body of the patient must be washed
with pure water after the disinfection to prevent allergic reactions.