Use Instructions

ELST026-AUM001-V0.6.2 Page 62 of 68
NOTE: U
T
is the a.c. mains voltage prior to application of the test level
Guidance and manufacturer’s declaration - electromagnetic immunity
The Acure device is intended for use in the electromagnetic environment specified
below. The customer or the user of a Acure device should assure that it is used in
such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
Guidance
Electromagnetic RF energy
emitting equipment should be used
no closer to any part of a device,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter. Recommended
separation distances d in meters
(m) where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer. Field
strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey
a
, should be less than the
compliance level in each frequency
range.
Interference may occur
in the vicinity of
equipment marked with
the following symbol:
Conducted RF
IEC 61000-4-6
AM
1 kHz/80%
3 V
rms
0.15-
80 MHz
6 V
rms
1.8-2.0,
3.5-4.0,
5.3-5.4,
6.765-6.795,
7.0-7.3,
10.10-10.15,
13.553-13.56
7, 14.0-14.2,
18.07-18.17,
21.0-21.4,
24.89-24.99,
26.957-
27.283,
28.0-29.7,
40.66-40.70,
10 V
rms
0.15-80 MHz
30 V
rms
26.957-27.283
MHz
d = 0.35√p 0.15-80 MHz
d = 0,12√p 26.957-27.283 MHz