Use Instructions

ELST026-AUM001-V0.6.2 Page 64 of 68
8.2 Technical safety inspection
device with serial number ............. is / is not
4
in good working order
Inspection performed by:
Owner:
Location:
Name:
Name:
Date:
Initials:
Initials:
4
Cross out what does not apply.
If a specific test does not apply to this equipment, place a mark in the NA (not applicable) column.
8.2.1 Test 1: General
Yes
No
NA
1.
The results of earlier safety inspections are available.
2.
The logbook is present.
3.
The type plate and the supplier’s label are legible.
4.
The housing, connectors and display are undamaged.
5.
The power supply and connection cord are undamaged.
7.
The electrode connectors and cables are undamaged.
8.
The cables and connector of the hand piece are undamaged.
9.
The hand piece does not have any cracks or other damage that could
compromise the insulation.
10.
The automatic self-test at switch-on does not give an error message.
See §2.7
11.
The display does not show any defective points or lines.