8.2.2 Test 2: Patient circuit
Yes
No
1.
Connect a load of 1000 Ω between the hand piece needle entry and the
1-pole patient cable. Connect an oscilloscope to the 1000 Ω load.
(Needle side is the negative pole)
2.
Select current in mA-level and adjust to maximum intensity.
3.
At maximum intensity, the output currents corresponds within 10% with the
values on the display.
4.
The output signals correspond with Figure 1.
5.
The warning Patient circuit is not closed is given if the load is disconnected.
Figure 1.
8.2.3 Test 3: handpiece
Yes
No
1.
Go to the system settings menu and select the item hand piece.
Select the item: check the hand piece buttons.
2.
Follow the on screen instructions.
Press one hand piece button at the time.
Is pressing of all 3 buttons on the hand piece correctly shown in the
hand piece picture?
3.
Go to the system settings menu and select the item hand piece.
Select the item: check the coloured illuminated ring
4.
Follow the on screen instructions.
Verify the illuminated ring has the same color as selected on the screen.