8.2.4 Test 4: Electrical safety test (IEC 62353)
Yes
No
1
The touch current is less than 1000 µA
2.
The patient leakage current is less than 5000 µA
Notes:
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
8.2.5 Used test equipment
Measurement
Equipment name
SN or identification
Calibration
due date
Safety test
Electrotherapy
Hand piece