Use Instructions

Operating the Device
23
EN
Transponder signal lossThe device stops and cannot be restarted if the transponder loses its signal dur-
ing current use. If the signal is restored with 20 seconds, it is possible to continue
using the tube set.
6.3 Usable Tube Sets Overview
The following table lists the function of each type of tube set depending on the
selected indication. Explanation: [x] = usable, [-] = not usable
Table 6-1
CAUTION!
The pressure chamber diaphragm is fragile and can be replaced if damaged. The
reusable tube sets must be prepared before each use. Instructions are listed in
chapter 13.6 Care of the Reusable Tube Set.
WARNING!
The vacuum tube sets for this device contain diethylhexylphthalate (DEHP),
which is classified as toxic to reproduction according to the EU Directive 1272/
2008/EEC on Classification, Labeling and Packaging of Dangerous Substances.
DEHP may impair fertility and may cause harm to the unborn child. Therefore,
this product must not be used for unauthorized applications. When applied
within the intended use, the potential risk to pregnant or breastfeeding women
as well as to children resulting from the DEHP contained in this product is not
critical. In regard to the short exposure time and the physical characteristics, the
eventuality of critical quantities of DEHP being dissolved from the tube sets is
negligible.
WARNING!
Sterilize reusable instruments and tubing before surgery to prevent infections.
Check all the single-use/disposable items before removing them from the pack-
age to ensure that the packaging is intact and that the expiration date is still val-
id.
WARNING!
Reprocessing of sterile disposable products
Infection hazard for patients and/or users and impairment of product function-
ality due to reuse. Risk of injury, illness or death due to contamination and/or
impaired functionality of the product! Do not reprocess the product.
Art. No. Tube category Tube type LAPAROSCOPY HYSTEROSCOPY
T0505-01 Irrigation tube set Disposable (one-time use) x x
T0506-01 Irrigation tube set Reusable (NOT FOR SALE IN
USA)
xx
T0504-01 Vacuum tube set, incl. filter Can be used 30 days x x
T0502-01 Suction tube set (1 connection) Disposable (one-time use) x x
T0503-01 Suction tube set (2 connections) Disposable (one-time use) x