Use Instructions

Using the Laparoscopy Indication
33
EN
8.3 Inserting a Tube Set
Laparoscopy tube typesThe following tube sets can be used with the laparoscopy indication (ordering in-
formation can be found in chapter 19 Accessory List):
Table 8-1
CAUTION!
When inserting a tube set already filled with irrigation fluid into the tube retain-
er, it is possible to damage the membranes. Only insert unfilled tube sets to
avoid such damage.
For additional information, see 6.2 Using the Tube Sets and 6.4 Irrigation tube set
connection.
8.4 Starting and Stopping Irrigation
During the laparoscopy indication, the pump works with a permanently preset
pressure. The nominal flow depends on the nominal pressure and may reach a
max. value of 2.0 l/min.
Starting the irrigation process1. Insert the tube set (see chapter 8.3 Inserting a Tube Set).
2. Open tube clamps at irrigation tube.
3. Open inflow valve at instrument.
4. Press the [START] function field to start the irrigation process. The roller
wheel starts to turn.
5. Wait until the tube set is completely filled with fluid.
6. Press the [STOP] function field. The roller wheel stops. The irrigation process
is canceled or ended.
7. Relieve any remaining pressure in the irrigation tube by opening the instru-
ment.
8.5 Opening and Configuring the Laparoscopy Indication Menu
Menu overviewDevice parameters are displayed and changed in the [LAPAROSCOPY] menu. The
following chart provides an overview over the possible settings.
Article no. Tube category Tube type
T0505-01 Irrigation tube set Disposable (one-time use)
T0506-01 Irrigation tube set Reusable (NOT FOR SALE IN USA)
T0504-01 Vacuum tube set, incl. filter Can be used 30 days
T0502-01 Suction tube set (1 connection) Disposable (one-time use)