Use Instructions

44
Using the Hysteroscopy Indication
EN
9.14 Opening and Configuring the Hysteroscopy Indication Menu
Menu overview Device parameters are displayed and changed in the [HYSTEROSCOPY] menu.
The following chart provides an overview over the possible settings.
1. Press the [Menu] symbol while in the hysteroscopy indication to access the
settings. Press [HYSTEROSCOPY].
2. Press one of the 4 function fields: [Deficit], [Fluid bag], [Scale test], [Instru-
ment recognition] or [Factory settings].
The following is a detailed description of how to change the indication-specific
device parameters.
9.14.1 Setting the Deficit Threshold
Set deficit threshold The deficit threshold defines the threshold value for triggering the warning mes-
sages.
1. Press the [Deficit] function field. The deficit threshold can be set within a
range of 0.1 to 2.5 l.
2. Press the º or » symbol to increase or decrease the deficit threshold. Pressing
this symbol briefly increases or decreases the value in increments of 0.1 l,
pressing it longer in increments of 0.2 l/min.