Use Instructions

Using the Hysteroscopy Indication
45
EN
3. Press the symbol to return to the previous menu level while saving the
values.
Default setting: 1.0 l
9.14.2 Fluid Bag Configuration
Configure warnings1. In the menu, press the [Fluid bag] function field.
2. Press the [ON] function field to enable or the [OFF] function field to disable
the bag change warning function. The active function field is depicted in
green.
3. Press the symbol to return to the previous menu level. The setting is au-
tomatically saved.
Default setting: Off
Adjusting the Height (Use without fluid
monitoring unit)
1. Press the º or » symbol to increase or decrease the height between device
and fluid bag. Pressing this symbol briefly increases or decreases the value in
increments of 0.1 m, pressing it longer in increments of 0.5 m.
2. Press the symbol to return to the previous menu level while saving the
values.
Default setting: 1.00 m
Set volume1. Press the º or » symbol to set the volume of the fluid bags to a value between
0.5 and 5 l.
2. Press the symbol to return to the previous menu level. The setting is au-
tomatically saved.
Default setting: 1 l
(see Fig. 6-3 Attaching the fluid bags)
NOTE!
The adjustment of the bag height is inactive when using a fluid monitoring unit
since the height of the bags is predetermined by the fluid monitoring unit.
9.14.3 Scales Test
See chapter 12.4
9.14.4 Setting the Instrument Recognition
1. In the menu, press the [Instrument recognition] function field.
2. Press the [ON] function field to enable, or the [OFF] function field to disable
the instrument recognition function. The active function field is depicted in
green.
3. Press the symbol to return to the previous menu level. The setting is au-
tomatically saved.
Factory setting: AN (responsivity)