Using the Hysteroscopy Indication
EN
9.14.5 Resetting to Factory Settings
Use the hysteroscopy menu to reset the hysteroscopy device parameters to the
factory settings.
Table 9-2
Reset to factory settings 1. Press the [Factory settings] function field to reset the HYS device parameters
to the factory settings.
You can now do the following:
2. Press the [OK] symbol to save the setting, or
3. Press to return to the previous menu level without saving.
9.15 Safety Functions
9.15.1 General Safety Functions
During device operation, the status line will display safety and warning messag-
es. These messages refer to the handling and safety of the device during use.
Please see chapter 11 Safety Functions for additional information concerning the
general safety functions.
9.15.2 Exceeding the Deficit Rate
Deficit rate exceeded If the deficit rate of 300 ml/min is exceeded, 5 warning beeps are emitted twice
and the screen displays the following in red letters: High Fluid Loss! The warning
message is repeated after 15 s.
9.15.3 Reaching and Exceeding the Deficit Threshold
The deficit threshold defines the threshold value for triggering the warning mes-
sages (see chapter 9.14.1 Setting the Deficit Threshold).
Differential volume = threshold value If the differential volume reaches the set threshold (0.1-2.5 l), 3 warning signals
(beeps) are emitted and the screen depicts the following: Deficit reached!
The deficit value on the screen is now displayed with a red progress bar.
Differential volume > threshold value If the deficit volume exceeds the set threshold value, 5 short warning beeps are
emitted and the screen depicts the following in red letters: Deficit exceeded!
Every additional 100 ml the value exceeds the threshold value is indicated with
the same warning signal.
Hysteroscopy device parame-
ters
Hysteroscopy factory settings
Nominal pressure 60 mmHg
Max. flow range (max. flow) 800 ml/min
Deficit limit 1.5 l
Fluid bag height 1.0 m
Fluid bag size 1 l
Fluid bag warning OFF