Use Instructions

62
Annual Inspection
EN
Hysteroscopy Indication:
Nominal pressure = 100 mmHg
Nominal flow = 500 ml/min
Laparoscopy indication:
•Flow rate low= 1000 ml/min
7. Press the [START] function field.
8. Wait until the tube set is completely filled with irrigation fluid.
Hysteroscopy indication:
Allow the device to pump fluid for at least 1 minute (instrument recognition
function).
9. Clamp off the instrument tube end in the measuring cup without stopping
the pump.
10. Empty the measuring cup and place the end of the tube back into the mea-
suring cup.
11. Release the tube end. The irrigation process runs. Press the start button of the
stopwatch.
12. Hysteroscopy indication:
After 2 minutes, press the [STOP] function field and the stop button of the
stopwatch at the same time. The measuring cup must contain approx. 1 l
(±10%) of fluid.
Laparoscopy indication
After 1 minute, press the [STOP] function field and the stop button of the
stopwatch at the same time. The measuring cup must contain approx. 1 l
(±10%) of fluid.
The basic function test has been successfully completed once these values have
been reached.
14.3 Pressure Measuring Test
The layout of the test and its setup are depicted in Fig. 14-1 Setup of pressure
measuring test.
The pressure measurement test checks the proper functioning of the pressure
measurement. This test requires a complete tube set and a container filled with
water. The height of the water column (hydrostatic pressure) is used for measur-
ing and is then converted to mm mercury column (mmHg). The height of the wa-
ter column above the pressure chamber has to match the value of the actual
pressure display after conversion.
Conversion formula: p (cm H2O) x 0.74 = p (mmHg)
1. Switch device on. Wait until the device check has finished.
2. Select one of the possible indications.
3. Press the [START] function field. Wait until the tube set is completely filled
and bubble-free.
Important: The tube set behind the roller wheel must be filled completely
with fluid.
Fig. 14-1 Setup of pressure measuring
test