Use Instructions

Technical Data Pump
71
EN
1
A tube set is not an application part in terms of the standard. However, it meets
all the technical requirements for an application part.
Measurement range
Throughput 0-2.0 l/min
0-300 mmHg
9995 ml
Pressure
Deficit/inflow volume
Precision (repeatability)
Flow ±10%
±10 mmHg
± 10% (relative to pumped inflow volume)
Pressure
Deficit
Accuracy
Flow ±10%
± 5% (of final value)
±10% (relative to pumped inflow volume)
Pressure
Deficit/inflow volume
Max. dimensions Width x Height x Depth 380 mm x 148 mm x 388 mm
Weight 6.3 kg
Interfaces:
IN/OUT signal for components 1x scale connection (D-sub connector, DE-9)
1x service connection (USB port, USB 2.0)
Mains power socket IEC 60320-1 C14
RFID Transponder Technology Working frequency: 13.5609 MHz,
Transmitting power: -7.51 dBμA/m at 10 m
Essential Performance Hysteroscopy
Pressure build-up in the body cavity, control and measure-
ment, limit value: max. 150 mmHg (normal condition),
First error: 200 mmHg for max. 5 seconds
Laparoscopy
Generate irrigation flow, 1.0 l/min, 1.5 l/min, 2.0 l/min, toler-
ance ±10%
First error: 0 l/min (no function) or 2.0 l/min (limit due to
design constraints)