User Manual User guide
English 8
Function test
After the components are removed from the package, do a function test.
1. Connect the connector cable to the sensor (8-pin polarized
connector) and an appropriate sc controller (5-pin polarized
connector) (refer to Connect sensor cable on page 8).
2. Apply power to the sc controller. The display is activated and the
sensor goes to measurement mode.
The sensor ticks quietly and regularly.
3. Cover the sensor measurement window with a sheet of white paper
(do not use recycled paper).
4. Vary the distance between the measurement window and the
paper. The measurement value on the display will change
accordingly.
Note: In air, the measurement value displayed is not exactly zero due to
reflections on the window surface (refer to Check the zero point on
page 14).
Installation
Connect sensor cable
1. Connect the polarized socket on the connector cable to the sensor
plug (8-pin plug) (refer to Figure 3 step 1).
2. Push the safety sleeve on the plug (step 2).
3. Hand-tighten the safety sleeve in position (step 3).
Figure 3 Connect the sensor cable to the sensor
CAUTION
Before power is applied, refer to the controller operation
instructions.
DANGER
Personal injury hazard. Only qualified personnel should conduct
the tasks described in this section of the manual.
NOTICE
If the sensor is not fully inserted, sun protection is recommended
in high ambient temperatures and intense solar radiation to
protect against thermal and UV effects.
CAUTION
Always put the cables and hoses in a position that does not bend
or cause a trip hazard.
CAUTION
Before power is applied, refer to the controller operation
instructions.