Instruction Manual

Manage devices
Install or remove input modules.
1. Select MANAGE DEVICES.
2. Select an option.
Option Description
SCAN FOR
DEVICES
The system shows the connected devices.
Note: If no device is connected, the system returns to
the main measurement screen.
DELETE DEVICE Removes the device when the device is no longer
connected.
Set up the calculation
Set up variables, parameters, units and formulas for the analyzer.
1. Select CALCULATION.
2. Select an option.
Option Description
SET VARIABLE X Selects the sensor referring to the variable X.
SET PARAMETER X Selects the parameter referring to the variable X.
SET VARIABLE Y Selects the sensor referring to the variable Y.
SET PARAMETER Y Selects the parameter referring to the variable Y.
SET FORMULA Selects the calculation formula to complete. Options:
None, X-Y, X+Y, X/Y, [X/Y]%, [X+Y]/2, X*Y, [X-Y]%/X
DISPLAY FORMAT Selects the number of decimal places shown in a
calculation result. Options: Auto, XXXXX, XXXX.X,
XXX.XX, XX.XXX, X.XXXX
SET UNITS Enters the unit name (maximum of 5 characters).
SET PARAMETER Enters the measurement name (maximum of
5 characters).
Configure the outputs
4–20 mA module setup
1. Select SETUP SYSTEM>SETUP OUTPUTS>4–20mA SETUP.
2. Select the OUTPUT.
3. Select an option.
Option Description
ACTIVATION The menu list items change with the selected function.
Refer to 4–20 mA activation options on page 12 for
more information.
SELECT
SOURCE
Select the output. Options: None if the output is not
configured, the analyzer name or calculation if a
calculation formula has been configured. Refer to Set up
the calculation on page 11.
SET
PARAMETER
Select the measurement channel from the list.
SET FUNCTION Select a function. Further options will vary depending on
which function is chosen. LINEAR CONTROL—Signal is
linearly dependent on the process value. PID
CONTROL—Signal works as a PID (Proportional,
Integral, Derivative) controller. LOGARITHMIC—Signal
is represented logarithmically within the process variable
range. BILINEAR—Signal is represented as two linear
segments within the process variable range.
SET TRANSFER If TRANSFER is or will be selected as the ERROR
HOLD MODE, select SET TRANSFER and enter the
transfer value. Range: 3.0 to 23.0 mA (default = 4.000).
Refer to Set the error hold mode on page 14.
SET FILTER Enter the filter value. This is a time-average filter value
of 0 to 120 seconds (default = 0).
SCALE 0mA/4mA Select the scale (0–20 mA or 4–20 mA).
English 11