Installation instructions DOC273.52.00146Jun06 Cleaning set LZY367 Introduction These installation instructions describe the installation and use of the cleaning set LZY367. Items supplied, cleaning set LZY367 Figure 1 Items supplied, cleaning unit LZY367 1 O-ring (7.5 x 2.5 NBR), 2 pieces 5 PharMed tubing 2 Cleaning adapter for FILTRATION PROBE sc, 2 pieces 6 Tubing (2 m long, 3.2 x 0.8), LZY195 3 Tubing (4 x 1, 0.7 m), 2 pieces 7 Set of fittings (3.
Cleaning set LZY367 Installation of the cleaning set LZY367 Danger! NEVER use 3% chlorine bleach together with 5% hydrochloric acid (Risk of the formation of chlorine gas). The chemicals used, in particular the chlorine bleach, are not allowed to be mixed or come into contact with other chemicals (Risk of the formation of chlorine gas). Take suitable protective measures for the chemicals used. If necessary wear personal safety equipment (e. g. gloves and safety glasses).
Cleaning set LZY367 Cleaning the FILTRATION PROBE sc 1. Ensure all steps in section Installation of the cleaning set LZY367 on page 2 have been performed. 2. Open the main menu on the analysis instrument. 3. Select the menu MAINTENANCE>TEST/MAINT>PREPUMPING>PREPUMPING PROBE and trigger a pump cycle. 4. Wait until the pump cycle is complete. Note: Ensure the tubes are not drained! 5. Hang the ends of the tubes on the cleaning adapters in a container filled with clean water and trigger a further pump cycle.
Cleaning set LZY367 Cleaning the analyser 1. Place the analysis instrument in the service state. a. For this purpose open the main menu on the analysis instrument. b. Select the menu MAINTENANCE>TEST/MAINT>SERVICE MODE and choose START. 2. Assemble the cleaning set as shown in the drawing Figure 6 Cleaning set assembled on page 7. 3. Remove the tube connection on the overflow vessel at the bottom right and fit the blind plug (see Figure 7 Fitting the blind plug and the cleaning set on page 8). 4.
Cleaning set LZY367 Figure 2 Removing the filter elements Figure 3 Fitting cleaning adapters 5
Cleaning set LZY367 Figure 4 Installation overview Figure 5 Installing the filter elements 6
Cleaning set LZY367 Figure 6 Cleaning set assembled 1 Disposable syringe (20 mL) 4 Union nut from set of fittings, LZY111 2 PharMed tubing 5 Tapered insert from set of fittings, LZY111 3 Tubing (2 m long, 3.2 x 0.
Cleaning set LZY367 Figure 7 Fitting the blind plug and the cleaning set Spare parts and accessories Description Tubing (2 m long, 3.2 x 0.8) Blind plug Fittings (3.
Warranty, liability and complaints HACH LANGE GmbH warrants that the product supplied is free of material and manufacturing defects and undertakes the obligation to repair or replace any defective parts at zero cost. The warranty period for instruments is 24 months. If a service contract is taken out within 6 months of purchase, the warranty period is extended to 60 months.
Contacts HACH LANGE LTD Pacific Way Salford Manchester, M50 1DL Tel. +44 (0)161 8 72 14 87 Fax +44 (0)161 8 48 73 24 info@hach-lange.co.uk www.hach-lange.co.uk HACH LANGE LTD DR. BRUNO LANGE GES. MBH Industriestrasse 12 A-3200 Obergrafendorf Tel. +43 (0) 2747 - 74 12 Fax +43 (0) 2747 - 42 18 info@hach-lange.at www.hach-lange.at DR. BRUNO LANGE AG HACH LANGE SA Juchstrasse 1 CH-8604 Hegnau Tel. +41 (0)44 - 9 45 66 10 Fax +41 (0)44 - 9 45 66 76 info@hach-lange.ch www.hach-lange.