Manual
21
Operation
CALIBRATE(continued)
CAL.CONFIG.>
VALUE CORR
Value correction (see section 4.6.3.5 on page 27)
Once value correction is complete, the correction data are displayed in the form of
the MATX2
1st NO3–N NO3DSC
Enter the NO
3
–N concentration (NO3D sc) of
the 1st point
1st CL- NO3DSC
Enter the Cl
-
concentration (NO3D sc) of the
1st point
1st NO3–N LAB
Enter the NO
3
–N concentration (lab) of the 1st
point
2nd NO3–N NO3DSC
Enter the NO
3
–N concentration (NO3D sc) of
the second point
2nd CL- NO3DSC
Enter the Cl
-
concentration (NO3D sc) of the
second point
2nd NO3–N LAB
Enter the NO
3
–N concentration (lab) of the
second point
-or-
CAL.CONFIG.>
PREVIOUS CAL
Selection of one of the last four matrix and value corrections performed
-or-
CAL.CONFIG.>
FACTORY CAL
Selection of the preset sensor code
DATE Date of last calibration
SENSOR CODE Enter the sensor code
CONFIGURE
EDIT NAME Entry or editing of the name. Up to 10 alphanumeric characters.
MEAS UNITS Selection of mg/L or ppm as measurement unit
SET PARAMETER Selection of NO
3
–N or NO
3
TEMP UNITS Selection of °C or °F as temperature unit.
TEMP CORR Entry of the temperature correction (–1.5 to +1.5 °C or –2.7 to +2.7 °F)
RESPONSE TIME Entry of the response time (30 sec to 300 sec)
LOG SETUP
Selects the interval for the data logging
(deactivate, 1, 2, 5, 10, 15, 30 min, 1, 2, 6 hr).
CL- COMPENSATE. Selection of the chloride compensation: On, off or fixed value
SET DEFAULTS Resets the configuration on the factory setting.
4.5 Sensor setup menu (continued)