Manual

46
Modbus Register
NO3 CONC 1 40079 Float 2 R 0/2576
NO3 mV CONC 1 40081 Float 2 R –250/400
NO3 mV drift
CONC 1
40083 Float 2 R –500/500
CL- CONC 1 40085 Float 2 R 0/2000
CL- mV CONC 1 40087 Float 2 R –300/400
CL- mV drift CONC 1 40089 Float 2 R –500/500
TEMP CONC 1 40091 Float 2 R 0/45
DATE 2 40093 Time2 2 R
NO3 N CONC 2 40095 Float 2 R 0/2000
NO3 CONC 2 40097 Float 2 R 0/2576
NO3 mV CONC 2 40099 Float 2 R –250/400
NO3 mV drift
CONC 2
40101 Float 2 R –500/500
CL- CONC 2 40103 Float 2 R 0/2000
CL- mV CONC 2 40105 Float 2 R –300/400
CL- mV drift CONC 2 40107 Float 2 R –500/500
TEMP CONC 2 40109 Float 2 R 0/45
OFFSET BY
NITRATE
40111 Float 2 R –70/50
SLOPE NITRATE 40113 Float 2 R 20/150
OFFSET BY
CHLORIDE
40115 Float 2 R –150/50
SLOPE CHLORIDE 40117 Float 2 R 20/100
NO3NmV 40129 Float 2 R –2500/2500
NitrateMeasmV 40131 Float 2 R –5000/5000
Nitrate VDrift 40133 Float 2 R –5000/5000 Drift in mg/L 5sec
Nitrate Noise 40135 Float 2 R –100/500 Noise in 10 seconds
CL-mV 40137 Float 2 R –5000/5000 Signal Cl-
ChlorideMeasmV 40139 Float 2 R –5000/5000
Chloride Drift mg/L 40141 Float 2 R –5000/5000 Drift in mg/L 5sec
Chloride Noise 40143 Float 2 R –100/+500 Noise in 10 seconds
pHDmV 40145 Float 2 R –5000/5000
Table 4 Sensor Modbus Register (continued)
Tag Name Register # Data Type Length R/W
Discrete
Range
Min/Max
Range
Description