DOC024.52.
Table of Contents Prerequisites ................................................................................................................3 Overview ........................................................................................................................3 Set up the link between server and client ....................................................... 3 OPC items ......................................................................................................................
Table of Contents 2
Prerequisites This document assumes the following is already installed on the 6110 instrument and the PC: 6110 Instrument • Operating system version 1.10 or later • TPA software version 2.11 or later Note: The above can be verified by selecting CONFIGURATION>Instrument parameters>System information from the instrument and checking the main board installed versions.
5. Start the OPC client application on the PC to display the initial screen. 6. Enter the CLSID number in the field followed by the instrument ID in the field. Click on the CLSID number and the server URL will automatically appear in the URL box. 7. Click on the Add Server button to establish the link between the 6110 instrument and the PC.
Note: If communications problems are encountered between the OPC client and server it may be necessary to adjust the DCOM and/or Firewall configurations. Information on these can be found on the internet or from your OPC client software supplier. 8. Click on the DA Browse tab to display the data available. Refer to the documentation for the OPC client software installed for details on accessing the data available.
All date fields are formatted according to the ISO 8601 extended format "YYYY-MM-DDThh:mm:ssZ" where "T" is the Date/Time separator and "Z" signifies "UTC time". An undefined date is sent as "0000-00-00T00:00:00Z" "Message.Out.Package" item The item "Package" is considered as Level 1 in the three level hierarchy described in OPC items on page 5.
Level 2 Level 3 CO2 P/T Max Min Enabled HS volume Max Min Enabled Liquid volume Max Min Enabled DO2 Max Min Enabled CO2 true Max Min Enabled Formula 1 Max Min Enabled Formula 2 Max Min Enabled Formula 3 Max Min Enabled Formula 4 Max Min Enabled Formula 5 Max Min Enabled "Message.Out.Measure" item The item "Measure" is considered as Level 1 in the three level hierarchy described in OPC items on page 5.
Level 2 Level 3 HS volume Liquid volume Package volume HS O2 quantity Liquid O2 quantity Total O2 quantity Liquid dissolved O2 concentration HS CO2 quantity Liquid CO2 quantity Total CO2 quantity CO2 pressure temperature CO2 true Liquid temperature Chamber temperature Barometric pressure HS pressure Formula 1 Formula 2 Formula 3 Formula 4 Formula 5 User value V1 V2 User comment C1 C2 C3 C4 C5 Mode Normalization Normalization temperature Warnings Errors 8 English
"Message.Out.Calibration" item The item "Calibration" is considered as Level 1 in the three level hierarchy described in OPC items on page 5.
Error value Error description 0x0080 No package 0x0100 Ultrasound 0x0200 Upper limit 0x0400 Clear flag detection 0x0800 P2 too low 0x1000 HS pressure unstable 0x2000 Abort 0x4000 Wrong computation error 0x8000 Unknown error "Measure.Warnings" item This item provides a list of any warning(s) associated with the measurement. This is a flag field where the indicated value is computed by the addition of all active warnings.
Note: The package must be created directly on the 6110 instrument, or through a web browser connected to the 6110 web server. It cannot be created using the OPC link "Messages.Message In.Show" item This item shows the specified message on the instrument. "Messages.Messages Out.State" item This output item indicates when the package can be changed for a new measurement analysis. State value Stet description Ready The package ID can be set in order to start a measurement analysis.
Error value Error description 0x00040000 Antifoam counter 0x00080000 Pending action 0x00100000 Pending manual action 0x00200000 No flow 0x00400000 RIM detector error 0x00800000 Upper limit 0x01000000 System initialization required 0x02000000 Communication error 0x04000000 ADIO board KO 0x08000000 CTRL board KO 0x10000000 EC board KO 0x20000000 CO2 foam out of specifications 0x40000000 Foam too many times "Messages.Message Out.
HACH COMPANY World Headquarters P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com © Hach Lange, 2013. All rights reserved. Printed in Europe. HACH LANGE GMBH Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info@hach-lange.de www.hach-lange.de HACH LANGE Sàrl 6, route de Compois 1222 Vésenaz SWITZERLAND Tel.