REFO 3/REFO 3-D Operating Instructions BDA 280 Edition 2
Operating Instructions REFO 3 / REFO 3-D Contents A PRODUCT DESCRIPTION A1 A2 A3 A4 A5 Concise Operating Instructions....................................... 2 Introduction....................................................................... 3 Scope of Delivery.............................................................. 3 Product Version ................................................................ 3 Instrument Set-Up.............................................................
Operating Instructions REFO 3 / REFO 3-D A Product Description A1 Concise Operating Instructions Calibration: 1. Put REFO on the working standard (remove cloth!) 2. Press key Meas/ON 3. Wait for display ok Gloss Measurement: 1. Turn on REFO with key Meas/ON 2. Use the program key to select program P1 3. Use the angle selector key to set the measuring angle 4. Place REFO on the sample to measure 5. Press key Meas/ON (measuring value is displayed) Difference Gloss Measurement: 1.
Operating Instructions REFO 3 / REFO 3-D Continual Measurement (only with REFO 3-D): 1. Turn on REFO with key Meas/ON 2. Select program P4 with the program key 3. Use the angle selector key to choose the first viewing angle to display, if required 4. Place REFO on the sample to measure 5. Start continual measurement by pressing Meas/ON key (3 measurements per second) 6. Move the REFO over the measurement sample and read off the relevant measurement value 7.
Operating Instructions REFO 3 / REFO 3-D 4. REFO 3 (for 115V) LMV 142 The functions regarding double calibration and sample memory described in the operating instructions apply to instrumention version REFO 3-D exclusively. A5 Instrument Set-Up Unpack the instrument carefully and check the delivery for missing parts (delivery note) or transport damages. Please keep the packing material for safe transport or shipment of the instrument.
Operating Instructions REFO 3 / REFO 3-D B Operation B1 Controls REFO 3-D is operated with the three blue keys next to and below the display and the large blue Meas/ON key on the instrument’s side.
Operating Instructions REFO 3 / REFO 3-D Meas./On-key Figure 2: REFO side view B2 Initial Start-Up B2.1 Charging the batteries The electrical power supply is provided by four Ni-Cd batteries, type LR6 AA Mignon. They are located in the battery compartment at the rear of the instrument. In order to charge the batteries • plug the charger plug into the mains/charging socket • connect the charger to the socket-outlet The batteries are loaded automatically within 14 hours.
Operating Instructions REFO 3 / REFO 3-D Battery Compartment mains / charging socket Figure 3: REFO rear view B2.2 Exchanging the batteries Note: When exchanging the batteries, all stored measurings will be lost if the cover of the battery compartment is lifted-off while the instrument is on. Therefore wait for the automatical shut-down before exchanging the batteries. To exchange the batteries: • Lift-off the cover of the battery compartment (see fig.
Operating Instructions REFO 3 / REFO 3-D 4. Press key Meas/ON to calibrate REFO. It identifies M- or D-mode automatically (not true for REFO 3) D 0379 Angle Select St - N r +1 → Program Key Statistics Key Figure 4: Entering the standard number 5. Enter the 4-digit standard number (current digit blinks). Press the statistics key until the correct digit appears. 6. Press the program key to skip to the second digit 7. Press the statistics key until the correct digit appears. 8.
Operating Instructions REFO 3 / REFO 3-D 2. Press the program key to skip to the second digit 3. Press the statistics key until the correct digit appears. 4. Repeat the procedure for the 3rd and 4th digits. 5. Press the angle selector key to skip to calibration value 60° 6. Input procedure as explained above 7. Press the angle selector key to change to calibration value 85° Input procedure as explained above Press the angle selector key to finish your input Now program P9 is finished automatically.
Operating Instructions REFO 3 / REFO 3-D B3 Configuration B3.1 Program Selection With the program key you can skip between the programs, i.e. every pressing the program key switches to the next program.
Operating Instructions REFO 3 / REFO 3-D C Applications C1 Calibration It is recommended to calibrate at least once a day or after extended breaks. Before starting to calibrate make sure that the standard is clean and the cloth removed.
Operating Instructions REFO 3 / REFO 3-D C4 Difference Gloss Measurement Difference measurement means measuring against a reference. The measured difference between reference and sample is the indicated measured value. • Turn on REFO with key Meas/ON • Use the program key to select program P2 • Use the angle selector key to select the viewing angle (the selected angle is indicated on the display) • Place REFO on the reference • Press key Meas/ON to measure the reference.
Operating Instructions REFO 3 / REFO 3-D C6 Statistics In all of the three programs P1, P2 and P3 you can acquire and evaluate entire setoff measurements with the following statistical values calculated: • mean value • standard deviation • coefficient of variability For statistical evaluation of a set-off measurement, procede as follows: • Turn on REFO with key Meas/ON • Use the program key to select program P1, P2 or P3 • Select the viewing angle for programs P1 and P2 with the angle selector key (for pro
Operating Instructions REFO 3 / REFO 3-D C7 Continual Gloss Measurement (REFO 3-D only) With the continual gloss measurement program, measurements are made automatically in 3 second intervals. In this measurement mode it is not possible to transfer data through the interface or the internal memory. This program function is enables the measurement of gloss differences over a large surface with the reflectometer.
Operating Instructions REFO 3 / REFO 3-D CAL: Calibrationvalues, standard number is printed, too Nr.: ordinary single- or three-angle measurement M: mean value S: standard deviation VK: coefficient of variability Nr. R: reference value Nr. d: difference measuring value Nr. dM: difference measuring value for statistics function dM: mean value of difference measurement dS: standard deviation of difference measurement Nr. RM: Nr.
Operating Instructions REFO 3 / REFO 3-D D Technical Information D1 Service Even a product such as the Dr. Lange REFO 3 will benefit from occasional care and attention. The outer casing can be cleaned using a dry cloth or mild household cleaner. Solvents such as benzine, acetone or similar substances must NOT be used. For maintenance of the calibration standard refer to chapter D.5. Cleaning of the optical parts should be carried out be our customer service. Further serviceing need not be carried out.
Operating Instructions REFO 3 / REFO 3-D Dr. Lange Nederland B.V. Laan van Westroijen 2a NL-4003 AZ Tiel Tel.: (0344) 631130 Fax: (0344) 631150 D4 Interface REFO 3 / 3-D has a serial interface installed to send the display data to an external device. Moreover, REFO 3 / 3-D is suitable for remote control with an external device.