Sigma 900 MAX Refrigerated Sampler Instrument Manual

Page 127
8990apdx_tbl.fm Troubleshooting and Error Messages
Appendix C Troubleshooting and Error Messages
Error Messages
Error Message Reason
*****Warning!*****
Logged data will be lost if you choose to continue? Yes/No.
Making a program change when a program is halted.
A minimum of 4 points are required.
The entered level value must be greater than the
previous value.
No values have been entered in the selectable table.
When entering level-flow or area-level tables, or when
attempting to use the tables, a minimum of four points must
be entered to create a valid table, and level values must
increase value as they are entered.
Bottle number must be: 1–xx
(where xx = #bottles in the tray).
When selecting a bottle to move to after selecting distributor
movement from the manual menu screen.
Bottle quantity must be: 1, 2, 4, 8, 12, 24. An invalid bottle quantity was entered.
Calibration requires a minimum submersion of six inches. Level sensor calibration of the bubbler or submerged sensor.
Day must be: 01—xx (where xx is the last day in month). Invalid number when setting a date.
Distributor movement halted due to key press.
A key was pressed during the distributor test (diagnostics) or
distributor movement from the manual menu screen.
Error: Distributor arm is not functioning properly!
The distributor test failed due to an arm obstruction, cable
not installed properly, or defect in the distributor assembly.
Function disabled while a program is running.
Use RUN/STOP key.
Attempted to use manual menu functions or the distributor
test while the program is running.
Hours must be: 00—23. Invalid number entered when setting a time, 24-hour format.
Hours must be: 01—12. Invalid number entered when setting a time, 12-hour format.
Incorrect password! Access denied. Incorrect password was entered (password is 9000).
Interval must be: 000:01–xxx:xx
(where xxx:xx is max. interval).
An invalid duration (hrs:min) was entered.
Invalid floating point format, only one decimal point allowed. A number was entered with two or more decimal points.
Logging intervals must be: 1,2,3,5,6,10,12,15,20,30,60. An invalid logging interval was entered.
Minutes must be: 00—59. Invalid number entered when setting a time.
No changes allowed while a pgm is running. Use run/stop key.
An inaccessible selection was made on a menu screen. Stop
the program before proceeding.
No retries are possible when the liquid sensor is disabled.
Occurs if Sample Retries is selected from the Modify
Selected Items scrolling list, and liquid sensor are disabled.
Enable liquid sensor to use retries.
No year entered. Invalid number entered when setting a date.
pH calibration failed—gain and/or offset out of range.
Try again.
The pH calibration calculations yielded an invalid number.
Try fresh buffers or a new probe.
Pumping took too long, please try again.
Occurs when the maximum time for timed calibration of a
sample volume (or rinse) is exceeded (the maximum is
65000 tens-of-milliseconds, which is roughly equal to
10.5 minutes).
Sampling halted because of key press.
Sampling halted because the bottle is full.
Sampling halted due to fluid presence after pre-purge.
Sampling halted during rinse—no liquid detected.
Sampling halted due to fluid presence after rinse purge.
Sampling halted—no liquid detected during sample phase.
May occur at the end of a sample attempt. Messages are
self explanatory. Investigate the tubing setup from strainer to
liquid sensor. Strainer may be plugged or out of water.
Tubing may not be installed properly in liquid sensor.
Sensor output voltage too low—cannot calibrate.
The D.O. calibration yielded an invalid result.
Try another probe.
Setpoint Sampling and Storm Water cannot both be enabled.
Enabled Setpoint Sampling with Storm Water enabled, or
enabled Storm Water with Setpoint Sampling enabled.