Sigma 900 MAX Refrigerated Sampler Instrument Manual

Appendix D
Page 135
8990apdx_pulsecounts.fm How to Calculate Pulses/Counts
Use the following examples to help you determine the value that you need to
enter when programming for sampling based on counts.
Example 1
You want to collect 24 samples over a 24-hour period on a 4–20 mA flow
proportional basis.
Average Flow Rate = 3.5 mgd (average flow rate over the 24-hour period)
Maximum Flow Rate = 10 mgd
a = 1440 minutes
1. Calculate Q.
2. Calculate t.
3. Multiply Q x t x 10.
Therefore, the value entered for
INTV = _ _ _ _ CNTS is 210.
Example 2
You want to collect 48 samples over a 24-hour period on a 4–20 mA flow
proportional basis.
Average Flow Rate = 1.75 mgd (average flow rate over the 24-hour period)
Maximum Flow Rate = 3 mgd
a = 1440 minutes
1. Calculate Q.
2. Calculate t.
3. Multiply Q x t x 10.
Therefore, the value entered for
INTV = _ _ _ _ CNTS is 175.
Example 3
You want to collect 96 samples over a 24-hour period on a 4–20 mA flow
proportional basis.
Average Flow Rate = 0.52 mgd (average flow rate over the 24-hour period)
Maximum Flow Rate = 2 mgd
a = 1440 minutes
1. Calculate Q.
Q
Average Flow Rate
Maximum Flow Rate
------------------------------------------------------
3.5 mgd
10 mgd
--------------------- 0.35 mgd===
t
a
n
---
1440 min
24 samples
------------------------------ 60 min/sample== =
0.35 mgd 60 min/sample× 10× 210=
Q
Average Flow Rate
Maximum Flow Rate
------------------------------------------------------
1.75 mgd
3 mgd
------------------------ - 0.583 mgd===
t
a
n
---
1440 min
48 samples
------------------------------ 30 min/sample== =
0.583 mgd 30 min/sample× 10× 174.9=
Q
Average Flow Rate
Maximum Flow Rate
------------------------------------------------------
0.52 mgd
2.0 mgd
------------------------ - 0.26 mgd===