Sigma 900 MAX Refrigerated Sampler Instrument Manual

Section 3
Page 59
8990startup.fm Advanced Sampling
18-E. Enter the desired high or low trigger point using the numeric keypad.
Press
ACCEPT to continue.
18-F. Enter the Deadband value or, if programming for Flow Rate Of Change
or Rainfall, enter a time interval when the flow or rainfall change must
take place (refer to Alarm Relays Programming on page 91).
18-G. Enter the number of bottles (out of the total set) to set aside for upset
samples. These will be the last bottles in the tray.
18-H. Select the method of distribution, Sample Per Bottle or
Bottles Per Sample.
18-I. Enter the sample volume using the numeric keypad.
18-J. Press
ACCEPT to continue to Variable Intervals.
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING:
HIGH TRIGGER POINT:
CANCEL
00000 in. CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING
DEADBAND:
CANCEL
0.000 CLEAR
ENTRY
(USE NUMERIC KEYPAD)
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING:
NUMBER OF BOTTLES
CANCEL
2CLEAR
ENTRY
ENTER 1 — 4
11:00 AM 21 - APR - 01 UPSET SAMPLES
ACCEPT UPSET SAMPLING
SAMPLE VOLUME:
CANCEL
0000 CLEAR
ENTRY
ENTER 10 — 9999