Sigma 900 MAX Refrigerated Sampler Instrument Manual

Page 69
8990options.fm Optional Device Installation
Section 5 Optional Device Installation
This section describes how to setup a rain gauge to the Sampler as well as
how to connect, program, calibrate, and maintain the optional water
quality probes:
5.1 Rain Gauge
An external “tipping bucket” rain gauge (Cat. No. 2149) can be connected to
the Rain Gauge connector of the sampler (Figure 21). The rain gauge
provides a dry contact closure to the sampler.
As rainfall collects in the 20 cm (8 in.) diameter funnel, it is directed into one
side of a “tipping bucket” assembly. As each bucket fills, it causes the bucket
assembly to tip and empty out into the bottom of the rain gauge. Each tip of
the bucket causes a single contact closure to the rain gauge and sends a
short 12 V dc pulse into pin C of the Rain Gauge connector. Each pulse (tip)
represents 0.025 cm (0.01 in.) of rain.
Figure 21 Rain Gauge Tipping Bucket
Rain Gauge (section 5.1 on
page 69
Dissolved Oxygen Probe
(section 5.4 on page 74)
pH Probe
(section 5.2 on page 70)
Conductivity Probe
(section 5.5 on page 76)
ORP Probe
(section 5.3 on page 72)
Table 7 Rain Gauge Base Board Connections (J5)
Pin Signal Description
A +12 V dc source output
B not used
C +12 V dc pulse input
D not used
E not used
F not used