Modem
8990cm.fm
Section 6
6.2.2.5 Cellular Modem Triggering
When Cellular Modem Triggering is enabled both the cell phone and modem
remains on for a user specified duration after calling a pager or remote
computer when an alarm condition is met. It is useful when calling back to a
site via modem, after receiving an alarm, or viewing the current status.
If Cellular Modem Triggering is disabled the sampler is still able to call a pager
or remote computer when an alarm condition is met. However, both the
modem and cell phone will immediately turn off after the call is completed.
Cellular Modem Triggered Duration
After enabling Cellular Modem Triggering, enter a value for Cellular Modem
Triggered Duration. This is the amount of time that the modem remains on
after it dials a pager or remote computer when an alarm condition has been
met. The value is in minutes and can be anywhere between 1 to 120 minutes.
Note: This display will not appear unless Cellular Modem Triggering is enabled.
6.2.2.6 Pager Option
The sampler can be set to call up to three individual pagers or a remote
computer when an alarm condition has been met. The pager setup is an
extension of the Modem Setup menus. To have the sampler call a pager, the
Pager Option must be enabled.
1. Enable the Pager Option, using the
CHANGE CHOICE key. Press ACCEPT.
11:00 AM 21 - APR - 01 MODEM SETUP
ACCEPT CELLULAR MODEM
TRIGGERING:
CANCEL
ENABLED CLEAR
ENTRY
CHOICES: ENABLED, DISABLED
11:00 AM 21 - APR - 01 MODEM SETUP
ACCEPT CELLULAR MODEM
TRIGGERING DURATION:
CANCEL
30 min. CLEAR
ENTRY
ENTER 1 - 120
11:00 AM 21 - APR - 01 MODEM SETUP
ACCEPT
CHANGE
CHOICE
PAGER OPTION:
CANCEL
ENABLED
CHOICES: ENABLED, DISABLED