Sigma 900 MAX Refrigerated Sampler Instrument Manual

Page 86
Modem
8990cm.fm
Section 6
Table 15 Pager Alarm Codes
Alarm Code Equipment Reason Alarm Code Equipment Reason
Low Main Battery 1 Battery pack is less than 11.5 V High CH5 28
Memory Battery 2 Internal memory battery is low High CH6 29
Low Slate Memory 3 Less than 10% slate memory left High CH7 30
Slate Memory Full 4 Slate memory is used up
High
Reference
Te mp.
31
Modem Failure 5 Modem chip/modem board failure
High
Velocity
32
Missed Sample 6 No liquid detected when sampling High D.O. 33
Purge Failure 7
Water present at sensors after
purge
High D.O.
Te mp.
34
Jammed Distributor 8
Sensors indicate arm is not
moving
High
Conductivity
35
Bottle is Full 9 Full bottle option indicator is on
High Cond.
Te mp.
36
U-Sonic Echo Loss 10 No return signal detected Low Level 37
Xducer Ringing 11
The return signal is detected too
soon
Low Flow 38
U-Sonic failure 12 Ultrasonic board detects an error
Low
pH/ORP
39
RS485 Timed Out 13 Comm. problems with RS485
Low
Process
Te mp.
40
Unable to Cool 14 AWRS only
Too high temperature in a
compartment
Low CH1 41
Unable to heat 15 AWRS only
Too cold a temperature in
compartment
Low CH2 42
Low Bubbler Pres. 16
(not
applicable)
Possible leak in bubble tank Low CH3 43
Clogged Bubbler 17
(not
applicable)
Bubbler tube is plugged Low CH4 44
High Level 18 Low CH5 45
High Flow 19 Low CH6 46
High Flow Rate of
Chg.
20 Low CH7 47
High pH/ORP 21
Low
Reference
Te mp.
48
High Process
Temperature
22 Low Velocity 49
High Rainfall 23 Low D.O. 50
High CH1 24
Low D.O.
Te mp.
51
High CH2 25
Low
Conductivity
52
High CH3 26
Low Cond.
Te mp.
53
High CH4 27