Manual Veloria Art. No. 0975 FB 04-267 Rev. 2 Date: 02.06.
Directory Foreword 3 Safety instructions Cleaning & Disinfection 4 7 Getting to know the unit 8 Before the use 9 Use Operation Working positioning Suction Changing filter bag Changing control unit filter Clamping the bur/ tool Setting up the speed Foot switch 11 12 14 14 14 15 15 15 Maintenance & Care Cleaning the chuck Maintance overview Handling instructions 16 17 17 Troubleshooting 19 Technical data Declaration of conformity Disposal Symbol Safety check Accessories 20 20 21 22 23 23 2
Foreword The Veloria foot care unit was produced according to strict quality criteria and uses the latest technology in its design. The images in this manual may vary slightly from the product you have bought as we are constantly developing our products. This user manual includes a detailed description and explanation of how to use the Veloria unit. If you have any questions or suggestions you can contact us by telephone or email. This user manual belongs with the Veloria unit. Keep it somewhere safe.
Safety Instructions This appliance can be used by children aged from 8 years and above and persons with reduced physical, sensory or mental capabilities or lack of experience and knowledge if they have been given supervision or instruction concerning use of the appliance in a safe way and understand the hazards involved. Children should be supervised not to play with the appliance. Cleaning and user maintenance shall not be made by children without supervision.
Safety Instructions Only when pulling the mains cable, you can ensure that the device will be completely disconnected from mains. Check the mains cable regularly for damages and in case of such, have it be changed or repaired by an authorized service center. Prevent your hair from wrapping itself around moving parts. If applicable, wear a hair net. When working with materials, which might create dust or moisture, use a drilling unit with suction or spray technique. Wear a nose and mouth protec-tion.
Safety Instructions Before using the equipment, assure that: the tool can be pushed far enough into the chuck (Test this with a dipstick and clean if necessary) the inserted tool sits properly in the chuck and will not cant/jam while clamping it the tool is approved for using it at the max speed (see manufacturer´s instructions), or only use the speed required by the instrument the tool is absolutely dry. For safety reasons do not change the tool while the motor is running.
Safety Instructions When the device is brought out of standby mode it will run with the rotation speed and suction power used last. Take care to not start the mill at an overly high rotation speed. Select the required rotation speed before changing the milling tool. Always follow the operating instructions. Do not carry out any repairs, modifications or maintenance work yourself. This is only to be accomplished by an authorized specialist.
Safety Instructions Cleaning and disinfection The device is designed to be compatible with rotating instruments which do not penetrate the skin or mucosal or inadvertently may still cause injuries. For this application please adhere to the cleaning and disinfection instructions provided below. After every treatment please carry out these steps: Remove the rotating tool (e.g. bur) first. To do so, pay attention to the instructions of its manufacturer.
Getting to know the unit 1 lernen 2 Gerät kennen 3 8 4 9 1 Decrease the suction output 2 Increase the suction output 3 Reduce the speed 4 Increase the speed 5 Display of speed/ suction level 6 Change direction of rotation 7 LED: ON during reverse rotation 8 Main switch 9 Switching ON/OFF the unit (Standby) 10 Connecting socket for foot switch 9 5 6 10 7
Before the use Please read through the instructions carefully before using the unit and carefully note the safety advice. Always keep this manual accessible to any users of the drill. Operation The working area must conform to the conditions described. Set up the unit in such a way, that the air slits are not blocked and the extracted air can escape well. To guarantee unhindered air flow there must a gap of at least 12cm to the right of the device.
Before the use Should the handpiece be twisted by the hose you can adjust that at the control unit. If you have a foot switch insert its plug into the port on the right of the device which displays this symbol . Check that the filter bag and dust filter are correctly fitted. Ensure that the tool is fixed. On the left side of the device you will find the main power switch. Press this switch to turn the unit on. To use the hand piece and suction function press the button (9) on the front of the unit.
Use Operation If you want to take a break or finish the job press the standby button (9). The handpiece motor and suction will turn off. The handpiece can now be placed in its holder on the right of the unit. Alternatively you can switch the device to standby by pressing and holding the button on the handpiece or by pressing the button on the foot switch. In the following section the order in which you should use the device is described. Additionally the individual steps will be explained in detail.
Use Work positioning The following images are examples of how to work on various parts of the foot. Always make sure that the foot is firmly gripped and that you can work freely and without strain.
Use 14
Use Suction Pull out the lid of the dust chamber and remove the old filter. Switch on the suction with button (2). The suction output can be regulated in 5 levels. You will see the selected level in the display. Level 0 would mean that the suction is turned off. Hold the filter in the middle to prevend it from bending and sliding off (see picture below). Therefore you avoid it from coming off again. In case the suction motor overheats, the unit will switch off automatically.
Use When the device is brought out of standby mode it will run with the rotation speed and suction power used last. Take care to not start the mill at an overly high rotation speed. Select the required rotation speed before changing the milling tool. Clamping the tool/ bur The handpiece is equipped with a chuck, which can hold all burs with a diameter of 2.35 mm (this is the diameter usually used for tools in podiatry). To open the chuck, push the knob forward with your thumb and leave it there.
Maintance & Care Set the screw spanner (wrench) on the axle and hold it tight. Now use the front of the second spanner to unscrew the chuck. Cleaning the chuck Because of the additional protection washer (see right), not much dirt is usually entering the handpiece. In the process of time, some dirt might collect itself within the handpiece, and might affect its functions. Therefore the chuck has to be cleaned monthly.
Maintance & Care origin (which was usually caused by the tools used in these circumstances).
Maintance & Care Working with high pressure will not improve the outcome of your work, since the motor will even be slowed down. A second disadvantage is the stronger impact on the ball bearings and the motor, which will shorten their life span in the long run. In case of sending the device by mail remove the dust filter before. Only suck up dust or milling waste, never suck up larger particles such as small bits of cotton wool as this may be get caught in the handpiece and lead to blockages.
Troubleshooting Handpiece vibrates, is very noisy and/ or becomes warm in the front part Use of bent burs/ tools Use a different bur/ tool to test this. The bur/tool itself can be tested with the hadewe Bur Shaft Tester (Art. No. (4990) to see if it is bent). Unit does not work at all. Check, if the main switch of the unit is turned on at the side of the unit.
Technical data 0975 Veloria Total weight: 2,7 kg Dimension: W273xH142xD186mm Input: 110V-230V, AC voltage 50Hz power consumption : 1A Fuse primary: F1: 3,15A time delayed safety fuse F3: 315 mA time delayed safety fuse Fuse Sec.: Thermal fuse Protection class II Accuracy of speed display: ±10% Handpiece: Type B applied part Diameter: 19-25 mm Length: 144mm Speed: 6.000 - 40.
Technical data Disposal The device and filter may contain infectious material such as residual sanding dust or skin particles. Therefore please bear in mind the following disposal instructions. Disposing of the device Old devices bought in Germany that correspond to the electronic device law should be sent directly to the manufacturer (hadewe). We will dispose of the device free of charge. The devices should not be taken to public disposal plants (WEEE Reg.Nr DE20392713, b2b device).
Technical data Symbol Caution! Important Note! Follow instructions for use Temperature limit (indicates the lower and upper temperature limit) Air pressure: range Air humidity: range Keep dry permissible permissible ON/ OFF button (standby) (no disconnection power supply) from main Change direction Suction Rotation Alternating current Direct current Safety fuse Type B applied part Socket foot switch Class II Date of manufacture 23
Technical data Safety Check A safety check must be done in compliance with the regulations of the country where the device is used. The inspection interval is once a year. Accessories Art. No. 5179 3752 0970 Micro filter filter Foot switch The clamping system is designed for rotating instruments, which have a shaft of 2,35mm in accordance with ISO EN ISO 179-1.
hadewe GmbH Grambartstraße 10 30165 Hannover Germany info@hadewe.de www.hadewe.